This historical checklist comes from Zhejiang Highpower. Its temperatures and maintenance intervals describe the source guidance, not universal requirements; the current instrument manual and site procedure take precedence.

  1. Calibrate before and after use. When the instrument is used frequently, schedule intermediate checks.
  2. Check the sampling system for leaks before measuring.
  3. Start the instrument in clean air. Connect the flue-gas sampling probe after the instrument stabilises.
  4. After a crash or power interruption, check for zero shifts and recalibrate on site with standard gas before resuming measurements.
  5. The source recommends running in clean air for another 5–10 minutes after measurement to protect the sensors.
  6. Replace filter elements and condenser filter pads regularly to prevent dust contamination and measurement errors.
  7. Keep the instrument away from heat sources and radiant heat. The source gives a sensor temperature of 22–25 °C; use the limits for the installed model.
  8. The source recommends charging portable instruments every 2–3 weeks. Follow the battery instructions for the model in use.
  9. For dusty, high-concentration flue gas, including de-NOₓ plant inlet/outlet and dust-collector inlet locations, provide suitable filtration.
  10. For low-concentration measurements, the source describes heating to 120–140 °C, dust removal and moisture removal. The appropriate conditioning method depends on the gas species and instrument.
  11. Handle the analyser carefully to prevent damage to internal components.