Tianlan Optoelectronics

TLG-1000

Dilution CEMS

A coordinated sampling, dilution and analysis system for low-concentration industrial emissions.

Measures
SO₂ · NOₓ · CO₂
Principle
Probe dilution with gas-specific analysis
Manufacturer
Tianlan Optoelectronics
Prepare an enquiry
Your project 0
TLG-1000 Dilution CEMS

Dilution at the sampling point

TLG-1000 dilutes a small stack-gas sample at the probe before gas analysis. The measurement chain brings the sampling probe, dilution control, gas-specific analysers and data acquisition together.

  • Controlled dilution reduces the moisture loading presented to the analysis system.
  • The supplied brochure describes a small critical-orifice sample flow and controlled dilution gas.
  • Calibration gas can enter at the probe to assess the complete measurement path.

Analysis and integration

The supplied system material combines Tianlan sampling and dilution equipment with gas-specific analysis, including Thermo Scientific 43i and 42i configurations. Third-party instruments retain their manufacturer identity.

Agree the concentration basis, moisture and oxygen correction, calibration method and site data interface as one system specification.

System characteristics

TLG-1000 — supplied dilution CEMS brochure
Core measurementsSO₂, NOₓ and CO₂ with gas-specific analysers
Raw sample flowApproximately 50 mL/min in the described configuration
Dilution ratio50–200:1; selected for the measurement duty
Calibration entry pointAt the probe for whole-path checks
Quality controlZero/span drift, system response, indication error checks and diagnostic recording

43i SO₂ analyser specifications

Analysis component — brochure PDF page 4
PrinciplePulsed fluorescence
Selectable spans0–0.05 ppm to 0–10 ppm; extended spans up to 0–100 ppm
Zero noise0.5 ppb RMS with 60-second averaging
Minimum detection limit1 ppb with 60-second averaging
Zero drift over 24 hours<1 ppb
Span drift over 24 hours±1% of full scale
Linearity±1% of full scale
CommunicationRS-485, RS-232 and Ethernet

42i NOₓ analyser specifications

Analysis component — brochure PDF page 4
PrincipleChemiluminescence; NO, NO₂ and NOₓ
Selectable spans0–0.05 ppm to 0–20 ppm; extended spans up to 0–100 ppm
Zero noise0.2 ppb RMS with 60-second averaging
Minimum detection limit0.5 ppb with 60-second averaging
Zero drift over 24 hours<0.4 ppb
Span drift over 24 hours±1% of full scale
Linearity±1% of full scale
CommunicationRS-485, RS-232 and Ethernet

TLG-100 CO₂ analyser specifications

Analysis component — brochure PDF page 4
PrincipleNon-dispersive infrared absorption
Measurement range0–25%
Repeatability≤1%, as listed in the brochure
Response time≤200 seconds
Zero driftWithin ±2% of full scale
Span driftWithin ±2% of full scale
Interference errorWithin ±2% of full scale
CommunicationRS-485, RS-232 and Ethernet

Technical enquiry

  • Ask the team for the current technical data sheet and application-specific configuration.