System / O₂ · CO · process gases
HN-LQL-X
Sample conditioning and gas analysis for challenging raw coal gas containing tar, dust and moisture.
Tianlan system integration / Thermo Scientific analyser
Extractive Ammonia Analysis
Bring ammonia-slip measurement into the SCR operating picture through extractive sampling and chemiluminescence analysis.

The supplied SCR presentation includes a Thermo Scientific 17i ammonia analyser within the monitoring system. It is a third-party analysis component, not a TURELINE-manufactured instrument.
The total response depends on the analyser settings, sampling line and conditioning arrangement. Specify response requirements for the complete sample path rather than describing the system as delay-free.
| Preset concentration spans | 0–0.05, 0–0.1, 0–0.2, 0–0.5, 0–1, 0–2, 0–5, 0–10 or 0–20 ppm |
|---|---|
| Zero noise | 0.50 ppb RMS with 60-second averaging |
| Listed minimum detection limit | 1.0 ppb with 60-second averaging |
| Response time | 120 seconds with 10-second averaging |
| Linearity | ±1% of span |
| Listed precision | ±0.4 ppb at a 500 ppb span |
| Zero drift over 24 hours | <1 ppb |
| Span drift over 24 hours | ±1% of full scale |
| Ambient operating temperature | 15–35 °C |
| Standard sample flow | 0.6 L/min |
| Ammonia conversion | A catalytic converter heated to 750 °C oxidises NH₃ to NO; NH₃ is derived from total converted nitrogen minus NOₓ |
|---|---|
| Described sampling configuration | Dilution probe with the total-nitrogen converter in the probe enclosure and a 17i analysis unit |
| Analogue output options | Voltage; optional isolated 0–20 or 4–20 mA |
| Digital interface options | RS-232, RS-485 and TCP/IP |
| Status outputs | Ten status relays and power-failure indication |
| Listed supply options | 100 V AC, 115 V AC or 220–240 V AC ±10% |
| Listed power consumption | 300 W analyser; 600 W converter |
Ask the team for the current technical data sheet and application-specific configuration.