Instrument / Sodium
TL4110Na
Trace sodium monitoring with a dedicated electrode arrangement and controlled sample alkalisation.
Tianlan Optoelectronics
Online Silica Analyser
Long-path colorimetric silica analysis for water-quality monitoring in power and industrial processes.

| Measurement ranges | 0–200 or 0–2,000 µg/L; custom ranges by agreement |
|---|---|
| Detection limit | <0.5 µg/L |
| Repeatability | ±0.5% of full scale |
| Accuracy | ±1% of full scale |
| Light source / optical path | 810 nm / 18 cm |
| Minimum measurement / response cycle | 12 minutes; measurement cycle is programmable |
| Channels | 1–6, according to the selected configuration |
| Sample pressure | 0.5 MPa, as listed in the manual |
| Sample flow | 50–1,000 mL/min |
| Ambient temperature | 5–40 °C |
| Power supply | 100–240 V AC, 50 Hz; 220 V AC nominal supply is also listed |
|---|---|
| Analogue output | 4–20 mA for each channel |
| Digital communication | Modbus RS-485 |
| Alarms | Programmable concentration, sample and reagent alarms for each channel |
| Mounting | Panel or wall mounting |
| Dimensions (W × H × D) | 480 × 750 × 350 mm |
| Measurement storage | 7,000 records |
| Listed reagent consumption | 3 litres per month |
| Calibration / cleaning | Automatic online calibration, manual multi-point calibration and automatic cleaning |
Ask the team for the current technical data sheet and application-specific configuration.