Instrument / Sodium
TL4110Na
Trace sodium monitoring with a dedicated electrode arrangement and controlled sample alkalisation.
Tianlan Optoelectronics
Online Phosphate Analyser
Online phosphate analysis with automatic calibration, reagent monitoring and a long optical path.

| Measurement ranges | 0–5, 0–10 or 0–20 mg/L; custom ranges by agreement |
|---|---|
| Detection limit | <0.05 mg/L for the low-phosphate configuration |
| Repeatability | ±0.5% of full scale |
| Accuracy | ±1% of full scale |
| Light source / optical path | 405 nm / 18 cm |
| Minimum measurement / response cycle | 7 minutes; measurement cycle is programmable |
| Channels | 1–6, according to the selected configuration |
| Sample pressure | 0.5 MPa, as listed in the manual |
| Sample flow | 50–1,000 mL/min |
| Ambient temperature | 5–40 °C |
| Power supply | 220 V AC, 50 Hz |
|---|---|
| Analogue output | 4–20 mA for each channel |
| Digital communication | Modbus RS-485 |
| Alarms | Programmable concentration, sample and reagent alarms for each channel |
| Mounting | Panel or wall mounting |
| Dimensions (W × H × D) | 480 × 750 × 350 mm |
| Measurement storage | 7,000 records |
| Listed reagent consumption | 3 litres per month |
| Calibration / cleaning | Automatic online calibration, manual multi-point calibration and automatic cleaning |
Ask the team for the current technical data sheet and application-specific configuration.