Measures
Silica
Principle
810 nm colorimetry with an 18 cm optical path
Manufacturer
Tianlan Optoelectronics
Prepare an enquiry
Your project 0
TL3110Si Online Silica Analyser

Long-path silica measurement

  • 810 nm light source with an 18 cm optical path.
  • Automatic calibration and programmable measurement cycles.
  • Reagent monitoring, automatic cleaning and stored measurement history.
  • Multi-channel sampling options.

Measurement specifications

TL3110Si — technical table in the supplied July 2024 manual
Measurement ranges0–200 or 0–2,000 µg/L; custom ranges by agreement
Detection limit<0.5 µg/L
Repeatability±0.5% of full scale
Accuracy±1% of full scale
Light source / optical path810 nm / 18 cm
Minimum measurement / response cycle12 minutes; measurement cycle is programmable
Channels1–6, according to the selected configuration
Sample pressure0.5 MPa, as listed in the manual
Sample flow50–1,000 mL/min
Ambient temperature5–40 °C

Installation and interfaces

TL3110Si — installation and operating data
Power supply100–240 V AC, 50 Hz; 220 V AC nominal supply is also listed
Analogue output4–20 mA for each channel
Digital communicationModbus RS-485
AlarmsProgrammable concentration, sample and reagent alarms for each channel
MountingPanel or wall mounting
Dimensions (W × H × D)480 × 750 × 350 mm
Measurement storage7,000 records
Listed reagent consumption3 litres per month
Calibration / cleaningAutomatic online calibration, manual multi-point calibration and automatic cleaning