Highpower — Integrated Environmental & Process Monitoring Solutions


0571-28930767

For a Cleaner, Healthier, Safer World

Articles
  • Articles
  • Products
Search
Details

5800-GC-FID Analyzer

Applications
Thermo Scientific 5800 online gas chromatography, using gas chromatography combined with flame ionization detector principles (GC-FID), not only allows immediate monitoring of total hydrocarbons (THC) and non-methane total hydrocarbons (NMHC) in the gas environment, but also for specific VoCs. The online gas chromatography monitor projects include: A non-methane total hydrocarbons (NMHC)/BBTEX companies, e.g., benzene, toluene, ethylbenzene, neighbor xylene, xylene, xylene, xylene, isopropene, styrene, etc./C customisation programmes such as ozone pre-exposed substances (PAMS) and special requirements for monitoring substances such as methanol and propol.

Thermo Scientific 5800 online gas chromatography, using gas chromatography combined with flame ionization detector principles (GC-FID), not only allows immediate monitoring of total hydrocarbons (THC) and non-methane total hydrocarbons (NMHC) in the gas environment, but also for specific VoCs.


i. The online gas chromatography monitorable projects include:


A non-methane total hidrocarbons (NMHC)


BBTEX companies, e.g. benzene, toluene, ethylbenzene, neighbor xylene, xylene, xylene, isopropene, styrene, etc.


C. Customization programmes, such as PAPS and special requirements monitoring substances such as methanol, propol


II. APPLICATIONS:


1. Refining and petrochemic industries


2. Electronic semiconductor industry


Lineboards


4. Third-party notary testing


5. Pharmaceutical manufacturing


6. Rubber/plastic products


7. Paints and inks


8. Automobile manufacturing and maintenance


9. Printing and packaging


10. Furniture manufacturing


11. Surface coating


12. Black metallurgical


III. Functional characteristics:


1. Customized analytical programmes, including software, hardware


2. Negative pressure sample design with instant/time performance check


3. Available current/historical trend maps


4. Individual calibration curves for individual analysts


5. Automatically switch the high and low range correction curve


6. Conformity with national standard methods


7. Measurable zei High 150℃ Heat samples.


8. Non-methane total hydrocarbons (NMHC) analysis within 1 minute


Technical parameters

Analysis methodology

Weather chromatography flame ion detection method

Measurement range

0–50 ppm / 500 ppm / 5000 ppm / 50000 ppm / 500000 ppm, as methane

Accuracy

±1% F.S. or 0.1 ppm, whichever is better

Test Limit

0.05ppm

Recurrence

±1% F.S. or 0.1 ppm, whichever is better

Zero Floating

±1%f.s/week

Float across points

±1%f.s/week

Analysis Time

Non-methane total hydrocarbons (NMHC) analysis within 1 minute

Sample flow speed

Samples for analysis :600ml/min

Sample update rate :3-10L/min. max.

Correction cycle

Daily/weekly or customized

System alarm.

Instrument failure/correction failure/detector (FID) failure

Output

Modbus/ Ethernet /RS232/4-20mA/RS485/USB

Sample temperature

15-150℃,85% RH( No Condensation )

Environmental conditions

Operating Temperature :0-45℃,0-85% RH( No Condensation )

Storage temperature :-20-60℃,0-85% RH( No Condensation )

Power requirements

AC220V, 50/60Hz,1.5KW

Dimensions

482mm(W)X567mm(D)X355mm(H)

Weight

45kg

 

System operating requirements:

Gas

FID fuel gas

H2: 30 ml/min, 2 bar; purity >99.999%; THC <0.1 ppm

FID-fired gas

No oil, no air. ,300ml/min,2bar,THC<0.1ppm

Air

No oil, no air. ,60ml/min,2bar,THC<0.1ppm

Zero gas correction gas

No oil, no air. , 2bar,THC<0.1ppm ( It's carrying air. )

No oil, no water, no nitrogen. , 2bar,THC<0.1ppm ( The gas is nitrogen. )

Transgression correction gas

Hydrocarbons at unknown concentrations are mixed in air (carrying air) with an inflow pressure of 1 bar

Hydrocarbons at unknown concentrations are mixed in nitrogen (carrying nitrogen) with an inflow pressure of 1 bar

Pipeways blow gas.

No oil, no air. ,10ml/min,2bar,THC<1ppm